Centrifuges
Due to improper operation, centrifuges have been the cause of many accidents. Use of centrifuges must meet the following 10 requirements:
- Centrifuges should only be used by qualified personnel who have read the manual and are familiar with the guideline for use of centrifuges
- Centrifuge must stop when the lid is opened
- Centrifuge must be secured against unbalance - loop contact must be able to stop the centrifuge
- The centrifuge in operation must be stable. The rotor must always be level. Loose parts must not be released during use
- Do not exceed the maximum speed specified on rotor equipment. The speed regulator must ensure this
- Centrifugation of hazardous substances or biological agents must take place in tightly closed centrifuge tubes
- Centrifuges must not be used in potentially explosive areas or for centrifuging large quantities of flammable liquid
- Only use packaging, tubes and bottles in accordance with the supplier's guidelines
- On some centrifuges the rotor must be re-tightened before each run
- All centrifuges must be inspected at least once a year by the supplier or other expert
Leave the centrifuge with a partially open lid to avoid condensation. The rotor and centrifuge chamber should be cleaned regularly.